Rev 1.1 VisualSonics Vevo 2100 Imaging System Operator Manual 415
Appendix C: Descriptions of control panel controls
In Color Doppler Mode and Power Doppler Mode: Applies to the color signal
data only. It does not apply to the B-Mode background data. Levels: Off, Low,
Med, High, Max. Helpful when you are studying abdominal organ tissue such as
liver, kidney and pancreas.
In Contrast Mode: Sets the process persistence filter level. Levels: None, MIP.
Physio Settings
When in a Mode window, activates the physiological settings panel.
Power
Activates Power Doppler Mode acquisition and begins displaying the power box
overlay over the B-Mode background image.
Pre Trigger
In Contrast Mode, starts an analysis based on the number of frames defined in the
General tab of the Preferences window.
Stores cine loop data for a predefined number of image frames acquired after you
press the control, as compared to
Cine Store
which stores data acquired before you
press the control. To ensure that the system stores your cine loop, select the Auto
SAVE at Scan Completion option in the General tab of the Preferences window.
In B-Mode, Color Mode, Power Doppler Mode, Contrast Mode: You define the
pretrigger's cine loop size in the Cine Loop Size section (page 71) of the General
tab in the Preferences window.
Presets
Active during image acquisition in all modes except 3D-Mode. This rocker switch
cycles you through all the preset groups of acquisition parameters for the active
imaging Mode. The list of presets include the transducer-specific presets as well
as any custom presets that other operators added to the system.
All presets are both mode dependent, transducer dependent and application
dependent.
Priority
Determines the threshold point on the gray scale above which the system does
not apply color data. The red marker along the left side of the gray scale indicates
the threshold point.
Push up to assign more priority to the color data. Pull down to assign less priority
to the color data and more priority to the threshold on the B-Mode grayscale bar.