346 VisualSonics Vevo 2100 Imaging System Operator Manual Rev 1.1
Chapter 47: Acquiring Contrast Mode images
Burst
Transmits an ultrasound pulse at maximum setting. This destroys the contrast agent in
the region of interest. In the cine loop the system displays a vertical green bar to mark the
destruction event.
Pre Trigger
In Contrast Mode, starts an analysis based on the number of frames defined in the
General tab of the Preferences window.
Stores cine loop data for a predefined number of image frames acquired after you press
the control, as compared to
Cine Store
which stores data acquired before you press the
control. To ensure that the system stores your cine loop, select the Auto SAVE at Scan
Completion option in the General tab of the Preferences window.
Image Sequence
In Contrast Mode this control starts a sequence of configurable events. When you press
the control:
1. The system begins to store image data for the predefined number of frames in the
cine loop, as configured in the Contrast Mode preferences (page 74) section of the
General tab in the Preferences window.
2. The destruction burst event (page 407) runs automatically:
Using a) the transducer that you connect to the front panel of the Vevo 2100
Imaging System, or using b) the external Vevo SoniGene transducer that you
connect to the Parallel port on the rear panel of the cart
At a predefined percentage point of the entire pretrigger cine loop length
For a predefined period in tenths of seconds between 0.1 and 1.0 seconds (defaults
to 0.5)
3. The system continues to acquire image data for the remainder of the predefined cine
loop size, but the image is not automatically stored when the loop is completed
unless you select Auto SAVE on Scan Completion for Contrast Mode in the General
tab of the Preferences window.
To configure the control for Contrast Mode:
In the Cine Loop Size section (page 71) of the General tab in the Preferences
window configure the size of the cine loop.
In the Contrast Mode preferences section (page 74) of the General tab in the
Preferences window configure the parameters for the destruction sequence.