230 VisualSonics Vevo 2100 Imaging System Operator Manual Rev 1.1
Chapter 35: Acquiring M-Mode images
Transmit Power
Adjusts the power of the ultrasound signal transmission.
Turn clockwise to increase power. Turn counterclockwise to decrease power. Between
1% and 10% power the control adjusts power in increments of 1%. Between 10% to 100%
power the control adjusts in increments of 10%.
Frequency
Adjusts the transmit frequency of the transducer between the higher and lower
frequency levels that are supported by the specific transducer. When you increase the
frequency you can improve detail at the focus depth but the system tends to lose detail at
deeper tissues.
Push forward to increase the frequency. Pull back to decrease the frequency.
Focal Zones
This control adjusts the number and configuration of focal zones on your B-Mode based
image.
Focal zones enhance the resolution across your image, while slightly reducing the
acquisition frame rate. The system always displays at least one focal zone, and you can
apply a maximum of two additional zones depending on the transducer. When you add
focal zones the system maximizes the resolution for a larger area of your image, and
reduces the acquisition frame rate.
To use this rocker switch control:
1. Push the rocker switch forward to cycle through the following focal zone application
sequence:
Single zone
Two zones, narrow
Two zone, wide
Three zones, narrow
Three zones, wide
2. Pull the rocker switch back to cycle back through the focal zone options in reverse.