360 VisualSonics Vevo 2100 Imaging System Operator Manual Rev 1.1
Chapter 48: Analyzing Contrast Mode images
Setting Description
Curve Fitting Calculates and plots a perfusion curve based on the following formula*:
, where:
y = Contrast signal (pixel intensity)
A = Peak of curve
B = Slope of the curve
C = Contrast signal offset
t = Time
t
0
= Time offset
To create the curve:
1. Click
Start Curve Fitting and select a data point on the graph at the
transition from the base line to the perfusion period.
2. Click a data point where the data begins to plateau and then click
Finish
Curve Fitting
.
The system calculates and plots the red perfusion curve.
3. Click
Export and export the data as an image or as a CSV file for further
analysis.
* Wei, 1998, Quantification of Myocardial Blood Flow With Ultrasound-Induced
Destruction of Microbubbles Administered as a Constant Venous Infusion
.
Chart Y Axis
Setting Description
Contrast
Select to plot the contrast intensity information from the contrast data, and to
make the Percent Area controls and the Calculation controls available.
B-Mode Select to plot the grayscale intensity data from the B-Mode image.
Auto Scale Select to view a system-calculated best-fit scale value.
Scale Max Type a scale value between 0-100,000.