VDO MS 2000X - Manual de Serviço Página 503

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 932
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 502
:SBUS<n> Commands 25
Keysight InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 503
:SBUS<n>:SPI:TRIGger:PATTern:MISO:DATA
(see page 792)
Command Syntax
:SBUS<n>:SPI:TRIGger:PATTern:MISO:DATA <string>
<string> ::= "nn...n" where n ::= {0 |1|X|$}
<string ::= "0xnn...n" where n ::= {0,..,9 | A,..,F | X | $}
The :SBUS<n>:SPI:TRIGger:PATTern:MISO:DATA command defines the SPI data
pattern resource according to the string parameter. This pattern, along with the
data width, control the data pattern searched for in the data stream.
If the string parameter starts with "0x", it is a hexadecimal string made up of
hexadecimal and X (don't care) characters; otherwise, it is a binary string made up
of 0, 1, and X (don't care) characters.
Query Syntax
:SBUS<n>:SPI:TRIGger:PATTern:MISO:DATA?
The :SBUS<n>:SPI:TRIGger:PATTern:MISO:DATA? query returns the current
settings of the specified SPI data pattern resource in the binary string format.
Return Format
<string><NL>
See Also "Introduction to :TRIGger Commands" on page 589
":SBUS<n>:SPI:TRIGger:PATTern:MISO:WIDTh" on page 504
":SBUS<n>:SPI:SOURce:MISO" on page 501
NOTE
The :SBUS<n>:SPI:TRIGger:PATTern:MISO:WIDTh should be set before
:SBUS<n>:SPI:TRIGger:PATTern:MISO:DATA.
Vista de página 502
1 2 ... 498 499 500 501 502 503 504 505 506 507 508 ... 931 932

Comentários a estes Manuais

Sem comentários